씰 Slide the battery compartment back into the monitor and press
gently inwards until it clicks home.
씰 After replacing the battery, check the performance of the
monitor as described in Chapter 6.
If you complete the replacement within about 1 minute, the date and
time settings will be retained. All of the other results will be retained
however long it takes to replace the battery.
Think of the environment. Dispose of used batteries sensibly.
Remove the battery if your Accu-Chek Active monitor is to remain
unused for some time.
Never throw batteries into a fire.
This is dangerous because they will explode.
7.2 Battery life and battery replacement
When testing blood glucose, pay close attention to the following
points. They affect the dependability of your Accu-Chek Active
monitor and the accuracy of your results.
Temperature range
씰 For performing blood glucose tests the permitted range is from
+10 °C to +40 °C.
씰 A test can be performed at the limits of the permitted range
(between +5 °C and +10 °C or +40 °C and +45 °C), but the
result should be interpreted with caution. A
is displayed.
씰 Tests cannot be performed if the temperature is below (under
+5 °C) or above (over +45 °C) these limits; a
TTT
message is
then displayed.
씰 Ensure your Accu-Chek Active monitor is stored between –40 °C
and +70 °C.
씰 The monitor will issue a message if it is too warm or too cold
to carry out a blood glucose test (see Chapter 8.2).