For CoaguChek XS System Knowledge Test
1. T
2. T
3. T
4. F – The sample must be applied to the test strip within 10 minutes of
removing the strip from the container.
5. T
6. T
7. F – If the built-in quality control does not pass, a flashing “QC” will appear on
the display.
8. T
9. T
10. F – The CoaguChek XS Meter stores up to 100 test results with time and date
11. The CoaguChek XS Meter can be cleaned with 10% bleach solution or 70%
isopropyl alcohol.
12. Error 3 indicates that the strip is expired. Ensure that the date is set correctly
on the meter. If it is incorrect, set the correct date. If the date is correct, turn
the meter off and remove the code chip and the test strip. Use the code chip
and a test strip from a new box of test strips.
13. Error 5 indicates an error applying blood to the test strip. It is caused by not
having a large enough drop of blood. When applying blood to the test strip,
massage your finger until you have a good blood drop and hold your finger
against the test strip until the meter beeps.
14. Confirm that the code flashing on the display matches the code on the
container of test strips you are using. Then press the M button to continue
the testing process.
15. After 15 seconds your blood may begin to clot, which could lead to an
incorrect result.