Roche Accu-chek Performa Instrukcja Użytkownika Strona 1

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Nie Roche Accu-chek Performa. Performa ® Meter Information - Vancouver Island Health Authority Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - Blood Glucose Meter Training

Blood Glucose Meter Training1

Strona 2 - Disclaimer

Accu-Chek Performa Operator Training © VIHA 2013 Coding a meter1. Make sure the meter is OFF2. Turn the meter over3. Remove the old code chi

Strona 3 - What is changing?

Accu-Chek Performa Operator Training © VIHA 2013 Coding a meter cont’d4. Turn the code chip over so the number faces down5. Push chip into t

Strona 4 - Learning Outcomes

Accu-Chek Performa Operator Training © VIHA 2013 Code Chip• Each box of test strips contains a code chip• Ensure this new chip is inserted i

Strona 5

Accu-Chek Performa Operator Training © VIHA 2013 Cleaning & Disinfecting the MeterClean the meter after each client test with a Percept

Strona 6 - • 5-second test time

Accu-Chek Performa Operator Training © VIHA 2013 Meter Cleaning Steps1. Put on clean gloves (and gown if required). 2. Make sure meter is of

Strona 7 - Arrow Buttons

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Test Strips15

Strona 8 - Changing the Battery

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Test Strip - Features• 0.6 µL sample size• Wipe 1stdrop and use 2nddrop•

Strona 9 - Changing the Battery cont’d

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Control Solution• Only use Accu-Chek Performa Control Solutions• Write t

Strona 10 - Coding a meter

Accu-Chek Performa Operator Training © VIHA 2013 Why do a Quality Control ?• Performing a Quality Control verifies the reliability of the te

Strona 11 - Coding a meter cont’d

Accu-Chek Performa Operator Training © VIHA 2013 When to do Quality ControlA Quality Control should be done:• Immediately prior to client te

Strona 12 - Code Chip

Accu-Chek Performa Operator Training © VIHA 2013 Disclaimer• This information is intended for use by Health Care Professionals • Patients wh

Strona 13 - Remember:

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – for Healthcare Professionals 1. Insert the test strip into the

Strona 14 - Meter Cleaning Steps

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d3. Select the control solution to be used4. Place the met

Strona 15

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d6. Touch the second drop to the front yellow edge of the

Strona 16 - • Wipe 1

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d– The QC result appears on the display, along with a cont

Strona 17

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d10. Result is in range if “OK” & the result alternate

Strona 18 - Why do a Quality Control ?

Accu-Chek Performa Operator Training © VIHA 2013 Troubleshooting - Quality Control is Out of RangeReview this list for possible causes:– Wer

Strona 19 - When to do Quality Control

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Blood Glucose Test 1. Insert a test strip, the meter turns on 2. Make sure the

Strona 20 - Professionals

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Blood Glucose Test – cont’d3. A test strip and flashing blood drop appear on t

Strona 21 - 4. Place the meter on

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Blood Glucose Test – cont’d5. Wipe the first drop away and touch the second dr

Strona 22 - (means that enough sample

Accu-Chek Performa Operator Training © VIHA 2013 Hi and Low Results Display• Hi Level - greater than 33.3 mmol/L• Low Level – less than 0.6

Strona 23

Accu-Chek Performa Operator Training © VIHA 2013 What is changing?Roche ACCU-CHEK® meters will be the meters used by all Health Authorities

Strona 24

Accu-Chek Performa Operator Training © VIHA 2013 Reference RangesVIHA Glucose Meter Normal Reference Interval 4.0 - 11.0 mmol/L(The meter wi

Strona 25

Accu-Chek Performa Operator Training © VIHA 2013 VIHA Critical Values(follow your facilities procedures for appropriate actions)Critical Hig

Strona 26 - 2. Make sure the code number

Accu-Chek Performa Operator Training © VIHA 2013 Documenting ResultsRemember - Always document the result as soon as possible to avoid error

Strona 27

Accu-Chek Performa Operator Training © VIHA 2013 Viewing Results - AveragesThe Perform meter has the ability to show 7, 14 or 30 day averag

Strona 28

Accu-Chek Performa Operator Training © VIHA 2013 Limitations of ACCU-CHEK®Test Strips (Refer to test strip package insert for further inform

Strona 29 - Hi and Low Results Display

Accu-Chek Performa Operator Training © VIHA 2013 Maintaining Competency Competency on the blood glucose meter is your professional responsi

Strona 30 - Reference Ranges

Accu-Chek Performa Operator Training © VIHA 2013 Learning OutcomesThese slides should help you understand: • Basic meter features • How to p

Strona 31 - Critical High Critical Low

Accu-Chek Performa Operator Training © VIHA 2013 Review of existing BGM practice/policy Client identification must be confirmed using a min

Strona 32 - Documenting Results

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Meter Features• 5-second test time• 0.6 – 33.3 mmol result range• Large

Strona 33 - Viewing Results - Averages

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Meter FeaturesDisplayRight and Left Arrow ButtonsOn/Off/Set ButtonBatter

Strona 34 - Key Points:

Accu-Chek Performa Operator Training © VIHA 2013 Changing the BatterySteps:1. Open the battery door on the back of the meter by pushing the

Strona 35 - Maintaining Competency

Accu-Chek Performa Operator Training © VIHA 2013 Changing the Battery cont’d93. Insert the new battery with the + side up.4. Put the battery

Komentarze do niniejszej Instrukcji