Display Shows Possible Cause(s) Possible Solution(s)
Communication
Warning
j Communication of bolus data
from the pump was not
successful. Therefore, pump
data are not available and the
bolus data may not be
accurate.
j Communication of bolus data
from the pump was not
successful. Therefore, pump
data are not available and the
active insulin amount may not
be accurate.
j Ensure the meter and pump have Bluetooth wireless
technology turned on.
j Ensure the meter and pump are within
communication range.
j If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.