General Product Information
142
Specication Meter Base unit Power supply unit
Battery voltage/type 3.7 volt rechargeable
(lithium technology)
N/A N/A
Input voltage +7.5 V DC +7.5 V DC 100 to 240 V AC
Input frequency DC DC 50 to 60 Hz
Input current 1.7 A (max) 1.7 A (max) 350 to 150 mA
Battery capacity 100 (subsequent)
measurements possible
after 360 min of charging
N/A N/A
Interfaces Charge contacts
IR port
Barcode scanner
RF card
(Channel 1-11 only)
Charge contacts
IR port
RJ45 Ethernet *
USB type B *
DC connector
Replaceable
AC input contacts
Data transfer rate RF card: up to 54 Mbps IR: 9.6K - 115K bps *
Ethernet: 10 Mbps *
USB: 12 Mbps *
N/A
Min. barcode width 4 mil (1 mil = 0.0254 mm) N/A N/A
CDRH/IEC CDRH Class I N/A N/A
Supported barcodes Code 128, Code 39,
Code93, EAN 13**,
Interleaved 2 of 5,
Codabar., GS1 DataBar
Limited
N/A N/A
* Not applicable to ACCU-CHEK Inform II Base Unit Light
**EAN 13 barcodes are not recommended for identication of operators and patients.
Specication Accessory box Code key reader
Height 3.35 in / 85 mm 0.72 in / 18.4 mm
Width 11.02 in / 280 mm 1.37 in / 34.8 mm
Length 10.71 in / 272 mm 2.78 in / 70.7 mm
Weight 1100 g 28 g
Storage temperature -13 to 158 °F
-25 to 70 °C
36 to 77 °F
2 to 25 °C
Relative humidity
(storage)
< 93% < 93%
Battery voltage/type N/A Not replaceable
Interfaces N/A IR port