Linearity Testing
85
In the Linearity Test menu, the levels available for the
lin earity test are displayed.
4 Touch L1 to start the subsequent test with this
(rst) level.
Inserting test strips
After selecting the level, you are prompted to insert the
test strip.
1 Remove the test strip from the test strip vial and
close the vial again with the cap.
2 Hold the test strip so the lettering “ACCU-CHEK” is
facing upward.
3 Slide the test strip into the test strip port as far as it
will go in the direction indicated by the arrows on
the test strip. The meter beeps.