Initial Startup
112
Options for Operator ID Entry
This menu allows you to specify if and how operator login
is performed.
1 Touch the desired option to enable it:
■ Operator Entry
– None: The meter can be used without operator
login.
– Prompt or Prompt (numeric only): Operator login is
required. The Operator ID can be entered manually
via keypad or via barcode scanner.
– Scan only: Operator login is required. Entry of ID via
barcode scanner only.
■ Operator Validation
– None: The operator ID can have any length.
– Length: The operator ID must be a specified length
(up to 20 characters). Enter the minimum and
maximum lengths after selecting this option.
– List: The operator ID must be on a list received from
the DMS.
– List with password: The operator ID must be on a list
received from the DMS. Additionally, a password
must be entered.
2 Touch to save the settings. If the Length
option is enabled, enter the desired numbers. Or:
3 Touch to exit this menu without saving any
changes.
The display returns in both cases to the Setup Menu.
Operator ID Entry
03/16/10
Operator Entry:
12:48pm
Prompt
None
Operator Validation:
Length
Scan only
None
Prompt (numeric only)
List
List with password
123
456
789
0
Op. ID Min Length 12:48
pm
03/16/10
8
123
456
789
0
Op. ID Max Length 12:48
pm
03/16/10
12