Appendix
145
Patient ID validation mode 0: none
1: length
2: list
3: list allowing entry if not
on list
4: length if numeric
0YY
Patient ID length
(Allowed characters: a-z, 0-9,
“.” [period], “-” [hyphen])
1-20 characters N/A N Y
Patient name length 0-25 characters N/A N N
QC algorithm
QC Algorithm 0: None (always OK)
1: Last result OK
2: Time of day (DMS only)
3: Shift (DMS only)
4: Hours
5: Strip count
6: Time of day rotate
(DMS only)
7: Shift rotate (DMS only)
8: Hours rotate (DMS
only)
9: Strip count rotate
(DMS only)
0 (Y) Y
QC hours (Number of hours between
required control measurements if QC
algorithm is set to “Hours” or “hours
rotate”)
0 – 9,999hours 24 Y Y
QC shift length (a PASS control remains
valid for twice this amount in time after
shift start if the QC algorithm is set to
“Shift” or “Shift rotate”)
1, 2, 3, 4, 6, 8, 12, 24 hours 8 N Y
QC shift start (time of the shift start if the
QC algorithm is set to “Shift” or “Shift
rotate”)
0 – 23 hours 6 N Y
QC strip count (number of tests a PASS
control remains valid)
1 – 999 50 Y Y
QC time of day 6 POCT1-A data fields
00:00 – 23:59
06:00
09:00
12:00
15:00
18:00
22:00
NY
QC time of day set 6 POCT1-A data fields
0 or 1
1 0 0 0 0 0 N Y
Subject/Attribute Range Default Device DMS