Maintenance and Care
121
Cleaning the meter
Wipe the surfaces with a soft cloth slightly
dampened (not wet) or you may spray the meter
per the directions mentioned below.
Remove the meter from the base unit prior to
cleaning.
Do not clean the meter while performing a patient
or control test.
Place the meter on a level surface while wiping over
the test strip port area, making sure that no liquid
enters the strip port.
Dry the meter thoroughly after cleaning. Visually
verify that no solution is seen anywhere on the
meter at the completion of cleaning. If cleaning
fluid is allowed to collect in any opening, severe
damage can occur to the meter and the base unit.
If using commercially available pre-moistened
cleaning cloths, squeeze off excess cleaning
solution or blot on a dry paper towel to remove any
excess cleaning solution before cleaning the
surface of the meter.
If spraying the meter, place it on a flat surface or
table. Wear gloves. Be aware that the meter may
become slippery when wet. Do not let the meter
drop! Do not spray while the meter is docked in the
base unit!
Cleaning the barcode scanner
window
The barcode scanner window should be cleaned periodi-
cally. Use a clean, dry cloth to wipe the barcode scanner
window.
NOTICE
Ensure meter/base unit (including connectors) are thor-
oughly dried after cleaning or disinfecting.
WARNING
Do not spray directly into the strip port!
Moisture in the strip port may lead to incorrect blood
glucose results. If you suspect that moisture may have
entered the strip port, perform a glucose control test.