Testing a Blood Sample
100
ST
AT tests
STAT tests are a limited number of tests that can be performed in emergency situations, see
page 76. If the meter is configured to perform STAT tests, you have the option of performing a
test by choosing STAT even though a QC Lockout is in place.
1 To perform the measurement without
performing a quality control test, touch
Patient Test:
2 Touch to confirm the number of
remaining STAT tests.
3 Perform the test.
When a STAT test is performed, this infor-
mation is stored with the result. The
number of STAT tests allowed is reduced
by 1. After all pending quality control tests
are performed, the specified number of
STAT tests is available again in case of a
new lockout.
The default setting is three for the number of STAT tests allowed when in QC Lockout.
Main Menu 09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Logout
QC!
STAT Info
3 STAT tests remaining
before lockout.
I-011
Test 09:15 am
04/19/2012
P-ID: PID01
Code: 184
2.5
INR
Smith, Henry
Op: Operator 1
Par: PT
STAT Test
21
%Q
04/19/2012 09:14 am