Roche CoaguChek XS System Instrukcja Operatora Strona 122

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 121
Extended Functionalities
122
Barcode scanner
Using a DMS
, the barcode scanner may be configured to accept only selected barcode symbol-
ogies (e.g., EAN-13 Barcodes are not allowed). Furthermore it is possible to define a barcode
mask for Patient ID and Operator ID. With masks being defined, the meter checks a barcode after
scanning and rejects it, if it does not match the mask.
Stored test results and comments
When performing a test, the test result will be stored along with additional information, including
the Patient ID, Operator ID, the kind of test performed, and optional Comments. The meter comes
with a default set of Comments that can be assigned to each test result. If you are working with
a DMS, these comments can be replaced by the comments in the DMS. The new wording will
then be available for selection from the Comments list in the meter.
The default preset comments in the meter are:
Up to 3 comments can be assigned to each result.
INR result comments QC result comments
Asymptomatic
Cleaned Meter
Doctor Notified
Sick
Travel
Lab Draw
No Action
Procedure Error
Under Medication
Will Repeat Test
Changed Dosing
Acceptable Control
Cleaned Meter
Doctor Notified
New Control Lot
New Lot Number
New Strip Lot
New Strip/Kit Lot
No Action
Procedure Error
Proficiency Test
Repeat Control Test
RN Notified
Switched QC Vial
Will Repeat Test
For more information and technical details, please refer to the manual of the Handheld Base
Unit and the Technical Note stored on the Handheld Base Unit itself.
Przeglądanie stron 121