Testing a Blood Sample
79
5 Testing a Blood Sample
What you need:
CoaguChek XS Pro meter
CoaguChek XS PT Test Strips and matching code chip
When testing with capillary blood:
Lancing device approved for professional use (e.g., the Accu-Chek Safe-T-Pro Plus
lancing device or the CoaguChek Lancets)
CoaguChek Capillary Tubes/Bulbs (optional)
Cotton ball and alcohol wipe
Important notes
Always
… close the strip container immediately after removing a test strip.
… operate the meter at an ambient temperature between 15 °C and 32 °C
(59 °F and 90 °F).
… place the meter on a level, stable surface (table) or hold it so it is roughly horizontal.
… follow the information on correct handling of test strips in the package insert.
… keep the test strip guide and housing clean. See also the chapter entitled Cleaning and
Disinfecting the Meter (starting on page 123).
Do not use glass capillary tubes or capillary tubes that contain anticoagulants.
Protection against infection:
When collecting samples always observe the general precautions and
guidelines relating to blood sampling (see page 14).
Dispose of all test strips used for patient testing in accordance with the
disposal policy of your laboratory or practice (see page 14)