Roche CoaguChek XS System Instrukcja Operatora Strona 111

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 110
Control Testing and Quality Control
111
14 If you want to add a comment, touch .
15 Select the desired predefined comment(s)
from the pick list (if configured) or
16 Touch Custom to enter your own custom
comment. Use the keypad (as with login)
to enter your comment. A comment may
be up to 20 characters in length.
17 Once you have selected the desired com-
ment(s), touch to return to the results
screen.
After the test result is displayed, touch
. You will be prompted to remove the
strip.
18 Remove the test strip from the meter.
If you are performing a 2-level control, you will
now be asked to proceed with the second
level.
19 Power the meter off.
20 Remove the quality control code chip from
the meter and store it with the controls.
21 Clean the meter if this becomes necessary
(see Chapter 9, Cleaning and Disinfecting
the Meter).
Dispose of controls and used test strips from control testing
in accord-
ance with the disposal policy of your facility
. The control solution con-
tains animal material, which should be considered as potentially
infectious.
QC Comment 09:15 am
04/19/2012
New Control Lot
Cleaned Meter
Doctor Notied
Acceptable Control
Custom
Remove 09:15 am
04/19/2012
Przeglądanie stron 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 155 156

Komentarze do niniejszej Instrukcji

Brak uwag